Ammonia Scrubbers
Prevent downstream analyzer damage by installing an ammonia scrubber before the condensing cooler.
Maxtec and Perma Pure are now united under Salaera. → Learn More


Prevent downstream analyzer damage by installing an ammonia scrubber before the condensing cooler.

How Ammonia Scrubbers Lower Your CEMS Operating and Maintenance Costs
NOTE: For Use With Cold Extractive, Condensing Cooler Based Sample Conditioning Systems. Ammonia Scrubbers are designed to remove ammonia from a gas stream to protect the analyzer from clogging due to the formation of ammonium salts. Ammonia is a highly reactive gas. It is sometimes added to stack gases to reduce the nitrogen oxide content of the gases by conversion to nitrogen and water (the DeNOx process).
When present in gas samples, ammonia will readily react with other components such as sulfur dioxide in the sample to form ammonium salts. This salt is relatively low-boiling, so it is present as a gas at the higher temperatures in the stack and in the heated line. When the sample cools down while passing through our dryer, it precipitates out as a solid, clogging the dryer or analyzer. For this reason, we recommend installing an ammonia scrubber before the condensing cooler to prevent damage to the analyzers downstream.
For Nafion™ tubing -based sample conditioning systems, Ammonia Scrubbers are required when any Ammonia might be present in the sample gas to protect the Nafion™ tubing dryers in the system, as Ammonia reacts with the membrane material and will lower the water transfer rate over time.
Our ammonia scrubber consists of a polysulfone or stainless steel shell housing filled with a phosphoric based scrubbing media and inert ceramic burl saddles. The water vapor in the sample activates the scrubbing media to produce phosphoric acid. The phosphoric acid then reacts with the ammonia in an acid-base neutralization reaction producing a phosphate of ammonia. This compound is a solid even at elevated temperatures, and deposits immediately within the ammonia scrubber as a visible salt residue.
For proper operation, the ammonia scrubber should be kept at a temperature above the sample dew point to avoid the loss of water-soluble analytes. For this reason, we now offers a heated ammonia scrubber which can be placed in the sample path.
| Model Number | Description | Maximum Presure | Maximum Temp | Ports |
| AS-200-08-PS | Polysulfone shell with stainless steel housing | 30 psig | 125° C | 1/4" FNPT |
| AS-200-08-PSH | Polysulfone shell with heater blanket | 30 psig | *See Note | 1/4" FNPT |
| AS-200-08-SS | All 316 stainless steel construction | 100 psig | 130° C (200° C with hex bolt replacing thumbscrew | 1/4" FNPT |
| AS-200-08-SSH | Stainless steel shell with heater blanket | 100 psig | *See Note | 1/4" FNPT |
*NOTE: The heated ammonia scrubber is supplied with 5 leads. It can be wired for 110VAC or 220VAC. You may use the provided thermocouple which will regulate the temperature to 90°C or you may wire the unit to your own temperature controller and readout. If using your own controller, use the max. temperatures for the unheated unit as a reference.
| Item # | Description |
| AS-200-3 | Replacement O-rings, 3 per set |
| AS-200-08-E | Replacement absorbent, burl saddles and reagent, 3 fillings |
| AS-200-08-EB | Replacement absorbent, burl saddles and reagent, 15 fillings |
| AS-200-08-EB OPT1 | Bulk supply of reagent only (no saddles), 15 fillings |
Replacing the Scrubbing Media:
To view more video content, visit our YouTube Page.