Solving Water Vapor Challenges in Total Organic Carbon Analysis with NDIR
What is TOC Analysis?
Total Organic Carbon (TOC) analysis measures the amount of carbon present in organic compounds in a sample of water or liquid. It is a non-specific indicator of purity and quality, often detected as CO2 in fractional ppm to ppb ranges.
Total organic carbon analyzers yield results that provide information on the safety of drinking water and water systems, product quality and giveaway, process optimization, and environmental impact in a variety of applications, such as:
- Municipal wastewater and treatment
- Airport runoff
- Industrial runoff
- Food & beverage
- Ultra pure water applications
- Drinking water
- Pharmaceutical manufacturing
- Semiconductor manufacturing
- Chemical manufacturing
- Power generation
- Environmental monitoring
Want to gain insights into your specific application?
Carbon can water sample can be measured by a nondispersive infrared (NDIR) detector or by a conductivity detector. When NDIR detection is used, carbon is measured in the gas phase. When conductivity detection is used, carbon is measured in the liquid phase.
NDIR detection is the primary detection method used, but this method faces challenges with water vapor interfering with analysis and maintenance.
What challenges does water vapor present to TOC analyzers using NDIR detection?
After a liquid sample is injected into the flow path of the analyzer it is oxidized one of three methods:
- UV persulfate oxidation
- Heated UV persulfate oxidation
- Combustion
Typically, combustion is used for dirtier samples with more contaminants and particulate. After the sample is oxidized to a gaseous state containing CO2, some water still remains in the sample. For the highest accuracy, sensitivity and reproducibility possible, both liquid water and water vapor need to be removed before the sample is carried to the NDIR detector by the carrier gas. This is because: